AT – db – 0050

Attrezzatura Nome Inventario n. Note/Info
VME 8010
Crate CAEN
x CAEN

CAEN VME 8010 Crate – Info

Features

  • Low cost compact solution
  • 19″ x 7U (6+1) enclosure
  • 21 slot for 6U x 160mm VME modules
  • 470W Power Supply
  • VME64 J1/J2 monolithic backplane
  • Short circuit protection
  • Over / Undervoltage protection
  • Over temperature protection
  • 1U space for Fan Unit
  • Powered by 100÷230 VAC, 50 ÷ 60 Hz
  • CBLT cycles supported

 

Technical Specifications

Mechanics

7U bin for 6U x 160 mm VME cards, 21 slots, 1U space for fan tray

Mains input

Auto range: 85 ÷264 Vac, 47÷63 Hz,
inrush current: <40 A @ 264 Vac
power. fact.: 0.99 Typical

Fuse

F10 A, fast type

Maximum Total Output Power

470 W @ 110 / 264 Vac

Output power

60 A @ +5 V, 6 A @ -12 V, 8.9 A @ +12 V

Isolation

CE acc. to EN 61010

Regulation

< 25mV for 0-100% load change

Noise and ripple

+5V < 50mV
+12V < 90mV
-12V < 90mV
(measured at 90% of full load)

RFI-rejection

CE EN 50081-1, FCC (emission), EN 50082-1 or 2 (immunity)

Over Current Protection

Trip off at 105-125% of the nominal values

Overvoltage protection

Trip off at 120-130% of the nominal values

DC Off (trip off)

within 5 ms if +5%, -2.5% deviation from adjusted nominal values, after overload, overheat, over voltage, under voltage (bad status) and fan fail trip off voltages and currents adjustable, processor controlled

Internal temperature limits

Cut off: 110 °C heat sink, 70 °C ambient, autom. maximum fan speed if air above VME modules >45°C

Operation

0÷50°C without derating

Temperature coefficient

< 0.2% / 10K

Stability

10mV or 0.1% within 24 hours 50mV or 1.0% within 6 months

Over Temperature Protection

Trip off all Output Voltages and PS internal fan

Efficiency

75% typical at 230Vac & 100% rated power

Cooling Airflow

408 m³/h (at maximum fan speed)